Screen Size: | 10.4′′ |
Ventilation Mode: | Ippv,A/C,Simv,Spont/CPAP,Sigh,Maun |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Ventilation modes: | IPPV; A/C; SIMV; SPONT/CPAP; MANU |
Ventilator parameter ranges | |
Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
Frequence(Freq) | 1bpm ~ 100 bpm |
I/E | 4:1~1:8 |
PEEP | 0cmH2O ~ 20 cmH2O |
Pressure triggering sensitivity(PTr) | -20 cmH2O ~ 0 cmH2O(Based on PEEP) |
SIGH | 0(off) 1/100 ~ 5/100 |
Apnea Ventilation | OFF,5 s ~ 60 s |
Monitored Parameters | |
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume(Vt) | 0 mL ~ 2000 mL |
MV Airway pressure |
0 L/min ~ 99 L/min 0cmH2O ~ 100cmH2O |
Oxygen concentration | 15 % ~ 100 % |
Graphical display: | |
P-T(pressure - time) | |
F-T(flow - time) | |
Size 1.Wooden case packing size(main engine):L 560*W 560*H 615mm , GW:35KG ; NW:17KG 2.Wooden case packing size(trolley):L650*W690*H 520 mm , GW:38KG NW:18G |
|
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <11.3±0.3V |
No tidal volume | No tidal volume within 6s |
High oxygen concentration alarm Low oxygen concentration alarm |
19%-100% 18%-99% |
High Airway pressure alarm Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm |
20cmH2O-100cmH2O 0cmH2O-20cmH2O 1L/min-99L/min 0-20L/min (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
Working conditions |
|
Gas source | O2, |
Pressure | 280kPa-600kPa |
Voltage | -220V±22V |
Power frequency | 50Hz±1Hz |
Input power | 40VA |