Customization: | Available |
---|---|
Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
Type: | Patient Monitor |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Transducer Type | Sidestream CO2 Sensor |
Sample Rate | 50 mL/min, ±10mL/min. |
Principle of Operation | Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts |
Initialization Time | Capnogram displayed in less than 4 seconds full specifications within 2 minutes |
CO2 Measure Range | 0 - 114 mmHg 0 - 15% 0 - 15.2kPa |
CO2 Resolution | 0.1 mm Hg 0 to 59 mm Hg 0.25 mm Hg 60 to 114 mm Hg |
CO2 Accuracy | 0 - 40mm ±2 mmHg 41 - 76 mmHg ±5% of reading 77 - 114 mmHg ±8% of reading Above 80 BPM ±12% of reading |
CO2 Stability | Drift over four hours shall not exceed 1 mmHg maximum Accuracy specification will be maintained over a 120-hour period |
Sampling Frequency | 100 Hz |
Respiration Rate Range | 2 to 150 Breaths Per Minute (BPM) |
Respiration Rate Accuracy | ±1 breath |
Compensations | Automatic Barometric pressure 400 mmHg to 800 mmHg Operator selectable O2, N2O, He and agent compensation |
Calibration | No routine user calibration required. |
Voltage Requirements | 5.0 VDC ±5% |
Power Consumption | Less than 1.2 Watts typical (Steady State) Up to 2 Watts maximum on power up (Warm up) |
Interconnection |
Standard Pin out:
|
Data Interface | RS232, bi-directional, 19200 baud rate, standard N-8-1. |
Data Output | CO2 gas concentration(mmHg), End-tidal CO2, Inspired CO2, Respiratory Rate. Gas and barometric pressure compensated when supplied by host. |
Temperature and Humidity | Operating: 0 to 45ºC, 10 to 90% RH, non-condensing Storage: -40 to 70ºC, <90% RH, non-condensing |
Water Resistance | IPX4 - Splash-proof (sensor head only) |